Title of article
Systematization of efficiency correction for gamma-ray disk sources with semiconductor detectors
Author/Authors
Chatani، نويسنده , , Hiroshi، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
11
From page
291
To page
301
Abstract
Full energy peak efficiency correction for disk sources has been systematically studied using the mapping method with two high-purity germanium detectors and two low-energy photon spectrometers. The following are found using only single-line (i.e., no coincidence summing loses) γ-rays: (1) The efficiency distributions on a plane parallel to the entrance window of semiconductor detectors is in perfect accord with Gaussian curves inside the circumference of the cylindrical Ge crystal, however, they deviate from the curves outside the circumference. (2) The width parameters of the Gaussian function fitted to the efficiency distributions have a systematic relationship with γ-ray energy. (3) The mapping method is of practical use and has satisfactory accuracy.
Keywords
Semiconductor detector , Efficiency correction , Efficiency distribution , Systematization , Mapping method , Disk source
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1999
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2180846
Link To Document