Title of article :
Study of position resolution with COPS
Author/Authors :
Moromisato، نويسنده , , J. and von Goeler، نويسنده , , E. and Reucroft، نويسنده , , S. and Eartly، نويسنده , , D. and Droege، نويسنده , , T. and Prokofiev، نويسنده , , D.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
We describe an optical beam sensing device that can measure transverse displacements with accuracy of a few microns. An analog readout prototype with a set of four linear CCD arrays arranged around a square aperture, and illuminated by a cross-hair laser, was built and tested by the authors. Performance tests results demonstrate a linearity better than 5 μm throughout an active transverse range of over 20 mm. The reproducibility of the measurements was better than 2 μm. Long-term stability is 5 μm.
Keywords :
CCD arrays , ALIGNMENT , Cross-hair laser
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A