Title of article :
On the calibration curve for determination of boron in tissue by quantitative neutron capture radiography
Author/Authors :
Alfassi، نويسنده , , Zeev B. and Probst، نويسنده , , Thomas U.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
Quantitative neutron capture radiography (QNCR) is an imaging method to visualize the boron contents in biological samples such as whole body cryosections of small mammalians as well as cell cultures. One of the main problems of QNCR is calibration. In this paper an equation for the calibration curve for determination of boron in tissue by QNCR is developed. A comparison of the experimental curves with those obtained by calculations are given.
Keywords :
QNCR , Calibration Curve , ?-Tracks , Boron determination , BNCT
Journal title :
Nuclear Instruments and Methods in Physics Research Section A