Title of article :
The RICH detector of the SELEX experiment
Author/Authors :
Engelfried، نويسنده , , J. and Filimonov، نويسنده , , I. and Kilmer، نويسنده , , J. and Kozhevnikov، نويسنده , , A. and Kubarovsky، نويسنده , , V. and Molchanov، نويسنده , , V. and Nemitkin، نويسنده , , A. and Ramberg، نويسنده , , E. and Rud، نويسنده , , V. and Stutte، نويسنده , , L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
4
From page :
149
To page :
152
Abstract :
A RICH detector was used during the 1996/7 fixed target run at Fermilab in experiment E781 — SELEX. The detector utilizes a matrix of 2848 phototubes for the photocathode. The figure of merit, N0, has been measured to be in excess of 100 cm−1. The ring radius resolution for multitrack events has been measured to be 1.8 mm, which gives K/π separation on a 2σ level up to a momentum of 165 GeV/c, and p/π separation up to 320 GeV/c. We will describe the design of the detector, discuss the stability during operation, and report on achieved results
Keywords :
Phototubes , RICH detector
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1999
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2182281
Link To Document :
بازگشت