Title of article :
Absolute calibration of photomultiplier based detectors – difficulties and uncertainties
Author/Authors :
Wright، نويسنده , , A.G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
6
From page :
507
To page :
512
Abstract :
Photomultiplier manufacturers can provide a calibration of quantum efficiency over a range of wavelengths with an accuracy of up to 2%. To convert these figures to absolute counting efficiency requires knowledge of photomultiplier collection efficiency, F. Traditional methods for determining F are discussed with emphasis on sources of error. Light sources emitting at a known photon rate allow the absolute quantum efficiency to be determined directly. It is important in all attempts at absolute calibration to appreciate the conditions which manufacturers apply when calibrating photomultipliers.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1999
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2182430
Link To Document :
بازگشت