Amendolia، نويسنده , , S.R and Bertolucci، نويسنده , , E and Bisogni، نويسنده , , M.G. and Bottigli، نويسنده , , U and Ciocci، نويسنده , , M.A and Conti، نويسنده , , M and Delogu، نويسنده , , P and Fantacci، نويسنده , , M.E and Maestro، نويسنده , , P and Marzulli، نويسنده , , V and Pernigotti، نويسنده , , E and Romeo، نويسنده , , N and Rosso، نويسنده , , V and Russo، نويسنده , , P and Stefanini، نويسنده , , A and Stumbo، نويسنده , , S، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
4
From page :
14
To page :
17
Abstract :
We have investigated the detection performance of GaAs detectors made with different thickness and contact geometries. A comparison is made between these detection capabilities and the imaging requirements for the following medical applications: digital mammography, digital chest radiography and nuclear medicine. Experimental results and preliminary images are presented and discussed.
Keywords :
X-rays , GaAs detectors , Medical imaging applications