Title of article :
Performance of superconducting transition edge bolometers for infrared photometry and X-ray micro-calorimetry
Author/Authors :
Hoevers، نويسنده , , H.F.C and Bento، نويسنده , , A.C and Bruijn، نويسنده , , M.P and Frericks، نويسنده , , M and Kiewiet، نويسنده , , F.B and Mels، نويسنده , , W.A and de Korte، نويسنده , , P.A.J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
5
From page :
247
To page :
251
Abstract :
The performance of voltage biased superconducting transition edge bolometers for infrared photometry and X-ray micro-calorimetry is described. Experimental results for the resistance–temperature curve of the superconducting phase thermometer, the current–voltage characteristic of the bolometer, and its current noise are presented and evaluated. Excess current noise turns out to be present when the bolometers are biased in the superconducting transition. The possible origin of this excess noise is shortly discussed. For the infrared bolometer an NEP of 2 ×10−17 W/Hz at 270 mK is obtained. The X-ray micro-calorimeter resolution is 60 eVFWHM at 155 mK for 5.9 keV X-rays, a factor 2.6 above the thermodynamical limit.
Keywords :
electrothermal feedback , Infrared detection , x-ray detection , Superconducting transition edge bolometer
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1999
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2183542
Link To Document :
بازگشت