• Title of article

    Counting losses due to saturation effects of scintillation counters at high count rates

  • Author/Authors

    Hashimoto، نويسنده , , Kengo and Yamada، نويسنده , , Sumasu; Hasegawa، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    9
  • From page
    502
  • To page
    510
  • Abstract
    The counting statistics of a scintillation counter, with a preamplifier saturated by an overloading input, are investigated. First, the formulae for the variance and the mean number of counts, accumulated within a given gating time, are derived by considering counting-loss effects originating from the saturation and a finite resolving time of the electronic circuit. Numerical examples based on the formulae indicate that the saturation makes a positive contribution to the variance-to-mean ratio and that the contribution increases with count rate. Next the ratios are measured under high count rates when the preamplifier saturation can be observed. By fitting the present formula to the measured data, the counting-loss parameters can be evaluated. Corrections based on the parameters are made for various count rates measured in a nuclear reactor. As a result of the corrections, the linearity between count rate and reactor power can be restored.
  • Keywords
    Preamplifier saturation , Counting loss , Dead time , Counting statistics , Scintillation counter , High count rate
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1999
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2184290