Title of article :
CdTe and CdZnTe detectors behavior in X-ray computed tomography conditions
Author/Authors :
Ricq، نويسنده , , Stéphane and Glasser، نويسنده , , Francis and Garcin، نويسنده , , Michel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
8
From page :
45
To page :
52
Abstract :
The application of CdTe and CdZnTe 2D array detectors for medical X-ray Computed Tomography (XCT) is investigated. Different metallic electrodes have been deposited on High-Pressure Bridgman Method CdZnTe and on Traveling Heater Method CdTe:Cl. These detectors are exposed to X-rays in the CT irradiation conditions and are characterized experimentally in current mode. Detectors performances such as sensitivity and response speed are studied. They are correlated with charge trapping and de-trapping. The trapped carrier space charges may influence the injection from the electrodes. This enables one to get information on the nature of the predominant levels involved. The performances achieved are encouraging: dynamic ranges higher than 4 decades and current decreases of 3 decades in 4 ms after X-ray beam cut-off are obtained. Nevertheless, these detectors are still limited by high trap densities responsible for the memory effect that makes them unsuitable for XCT.
Keywords :
Intense irradiation , Trapping/de-trapping , injection , CdTe/CdZnTe detectors , Space charge , X-ray computed tomography
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2000
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2185092