Performance of a microcalorimeter with a superconducting transition edge thermometer
Author/Authors
Hoevers، نويسنده , , H.F.C and Bento، نويسنده , , A.C and Bruijn، نويسنده , , M.P and Gottardi، نويسنده , , L and Korevaar، نويسنده , , M.A.N and Mels، نويسنده , , W.A and de Korte، نويسنده , , P.A.J، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
4
From page
192
To page
195
Abstract
The performance of a single-pixel microcalorimeter is described and analyzed. The microcalorimeter has a Cu absorber which is isolated from the heat bath by means of a SixNy membrane and is connected to a voltage-biased Ti/Au proximity bilayer thermometer with a transition temperature of about 150 mK. The measured FWHM energy resolution for 5.9 keV X-ray photons is 12.4 eV for this sensor while 2.36(kBT2C)1/2 equals 9.1 eV. The detector time constant is about 1 ms. The measured current–voltage characteristic and current noise spectra are analyzed in detail in the context of the energy resolution. High frequency excess noise is measured in the superconducting transition and is briefly discussed.
Keywords
Superconducting transition edge sensor , Microcalorimeter , electrothermal feedback , x-ray detection