Title of article :
An improved hand-held four-detector γ-probe for radioassisted oncological surgery
Author/Authors :
Dusi، نويسنده , , W and Angelotti، نويسنده , , P and Bollini، نويسنده , , D and Moroni، نويسنده , , C and Ricard، نويسنده , , M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
6
From page :
531
To page :
536
Abstract :
The performance of an improved intraoperative γ-probe for radioassisted oncological surgery is presented and discussed. The probe is based on a square array of four 5×5 mm2 coplanar CdTe room temperature semiconductor detectors and each detector has an independent read out electronic chain, allowing an original handling of the signal. Therefore, the search for γ-emission hot points may be carried out in two different, independent ways: akes the new probe sensitive to the bidimensional gradient of the γ-ray flux, measured on the scanned plane. Furthermore, the algebraic sign of the difference indicates in which direction the probe should be moved to best locate the γ-source. Due to these characteristics, both the errors in the localization of a single, small size γ-source and the average measuring time are practically halved with respect to those obtained with a conventional single detector probe. In addition, a computer analysis is performed in order to evaluate the performances of different types of collimators in reducing the contribution of a second source close to the source under examination. This situation occurs, for example, when the probe is used in the framework of the sentinel-node surgical technique.
Keywords :
X-ray probe , Gamma-probe , Surgical probe , X-ray detectors
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2000
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2186811
Link To Document :
بازگشت