Title of article
Accurate calculation of total efficiency of Ge well-type detectors suitable for efficiency calibration using common standard sources
Author/Authors
Sima، نويسنده , , O.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
11
From page
98
To page
108
Abstract
A new method for the calculation of coincidence summing out corrections appropriate for γ-spectrometry measurements in the Ge well-type detector geometry is proposed. The method is based on a simple formula for the total efficiency of the detector, using an exact analytical relation for the mean detector thickness, combined with approximate formulas describing the interactions in the source and in the materials interposed between the source and the detector. The method can be easily applied for establishing the corrected, nuclide independent, efficiency calibration curves of Ge well-type detectors using experimental calibration data obtained with the common mixed γ-ray standard source containing 60Co and 88Y.
Keywords
Well-type detectors , Total efficiency , Ge detectors , ?-ray spectrometry , True coincidence summing
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2000
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2187112
Link To Document