• Title of article

    Methods for precise photoelectron counting with photomultipliers

  • Author/Authors

    Dossi، نويسنده , , R. and Ianni، نويسنده , , A. and Ranucci، نويسنده , , G. and Smirnov، نويسنده , , O.Ju.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    15
  • From page
    623
  • To page
    637
  • Abstract
    A series of measurements has been performed on a THORN EMI 9351 phototube in order to investigate its response to a low light intensity. Precise procedures to determine the intensity of the incident photon flux have been developed and compared. The data show that the various approaches give consistent and reliable results, thus allowing the precise calibration of the device for applications of photon counting.
  • Keywords
    Photomultiplier , Photoelectron , scintillator , Calibration
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2000
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number