• Title of article

    CARTOGAM – a portable gamma camera for remote localisation of radioactive sources in nuclear facilities

  • Author/Authors

    Gal، نويسنده , , O and Izac، نويسنده , , C and Jean، نويسنده , , F and Lainé، نويسنده , , F and Lévêque، نويسنده , , C and Nguyen، نويسنده , , A، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    8
  • From page
    138
  • To page
    145
  • Abstract
    We have developed a compact gamma-imaging system, CARTOGAM, for remote localisation of radioactive sources in nuclear facilities. This system is under industrial development and commercialisation by the firm EURISYS Mesures. The most specific characteristics of CARTOGAM lie in its size (8 cm in diameter) and mass (15 kg for the detection head, including the shield), which make it portable by a person. As an example, CARTOGAM detects a 660 keV source producing a 0.4 μGy/h dose rate at the camera location in 10 min. The angular resolution at that energy ranges from 1° to 3°, depending on the field of view (30° or 50°) and scintillator thickness (2 or 4 mm). We present here a review of the specifications of the camera and show a few images illustrating its performance.
  • Keywords
    Gamma-ray imaging , Radiation imaging detectors , Radioactive contamination mapping
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2001
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number