Title of article :
Noise contributions to energy resolution in CsI(Tl) scintillation detector with PIN photodiode readout
Author/Authors :
Mikhailov، نويسنده , , M.A. and Panteleev، نويسنده , , L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
5
From page :
288
To page :
292
Abstract :
The different noise contributions to the energy resolution of a CsI(Tl)-photodiode gamma detector have been measured. A “light noise” originating from the crystal has been detected. It has been found that it could be the main restricting factor together with the photodiode leakage current. Temperature dependence of the “light noise” is reported.
Keywords :
scintillator , photodiode , Energy resolution , Noise
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2001
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2190763