Title of article :
A compact double crystal monochromator for electrochemistry beamline at PLS
Author/Authors :
Rah، نويسنده , , S and Kim، نويسنده , , S.N. and Kim، نويسنده , , G.H، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
4
From page :
388
To page :
391
Abstract :
A compact double crystal monochromator based on 16.5′′ CF flange has been designed, fabricated and installed for electrochemistry beamline at Pohang light source. The Bragg angle range of the monochromator is 7–75°. The mechanical design is modified from typical Boomerang design [J.A. Golovchenko et al., Rev. Sci. Instrum. 52 (1981) 509; J.P. Kirkland, Nucl. Instr. and Meth. A291 (1990) 185] to have fixed beam offset and single driving axis for spectroscopy experiments. The parallelism error of the crystals is minimized to less than 6 μrad for the range, by using a precision single axis linear guide, Also, the number of mechanical parts in the vacuum is minimized and 1.8×10−9Torr of vacuum is achieved without baking.
Keywords :
Modified Boomerang , Synchrotron , Double crystal monochromator , Beamline
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2001
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2191836
Link To Document :
بازگشت