• Title of article

    High-accuracy detector calibration at the PTB four-crystal monochromator beamline

  • Author/Authors

    Krumrey، نويسنده , , M. and Ulm، نويسنده , , G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    4
  • From page
    1175
  • To page
    1178
  • Abstract
    A four-crystal monochromator beamline has been installed by the Physikalisch-Technische Bundesanstalt at a bending magnet of the electron storage ring BESSY II. The monochromatic radiation with very high spectral purity and high spectral resolution in the photon energy range from 1.75 to 10 keV is used to calibrate detectors by comparison to a cryogenic electrical substitution radiometer as primary detector standard with relative uncertainties well below 1%. This is one order of magnitude better than all calibrations of non-energy-dispersive detectors in this spectral range previously performed. The beamline is also used for the characterization of optical components.
  • Keywords
    X-ray detector calibration , Four-crystal monochromator , Semiconductor photodiodes
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2001
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2192233