Title of article :
A wide-range direction neutron spectrometer
Author/Authors :
Luszik-Bhadra، نويسنده , , M. and DʹErrico، نويسنده , , Herwart F. and Hecker، نويسنده , , O. and Matzke، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
7
From page :
291
To page :
297
Abstract :
A new device is presented which has been developed for measuring the energy and direction of distribution of neutron fluence in fields of broad energy spectra (thermal to 100 MeV) and with a high background of photon, electron and muon radiation. The device was tested in reference fields with different energy and direction distributions of neutron fluence. The direction-integrated fluence spectra agree fairly well with reference spectra. In all cases, the ambient and personal dose equivalent values calculated from measured direction-differential spectra are within 35% of the reference values. Independent measurements of the directional dose equivalent were performed with a directional dose equivalent monitor based on superheated drop detectors.
Keywords :
Silicon detectors , Radionuclide sources , CERN , Direction spectrometry , superheated drop detectors , Unfolding
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2002
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2193474
Link To Document :
بازگشت