Title of article :
Reliability of using a fixed matrix in coregistration of combined PET–MRI in a split magnet design
Author/Authors :
Sawiak، نويسنده , , S.J. and Hawkes، نويسنده , , R.C. and Ansorge، نويسنده , , Peter R.E. and Carpenter، نويسنده , , T.A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
2
From page :
54
To page :
55
Abstract :
We consider the effects of using a fixed linear transformation to match positron emission tomography (PET) and magnetic resonance imaging (MRI) data acquired simultaneously using a split-magnet system. Estimates of the frequency offset in MRI scans were used to calculate geometric variability in MRI reconstruction as a consequence of mis-setting this parameter in addition to repeated estimation of the transformation matrix by manual measurements. None of the measured variability approached the resolution of the PET images, so we concluded that a fixed matrix can be reliably used in such a system.
Keywords :
Positron emission tomography , Magnetic Resonance Imaging , image registration
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2013
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2193567
Link To Document :
بازگشت