Title of article :
Structured light-based motion tracking in the limited view of an MR head coil
Author/Authors :
Erikshّj، نويسنده , , M. and Olesen، نويسنده , , O.V. and Conradsen، نويسنده , , K. and Hّjgaard، نويسنده , , L. and Larsen، نويسنده , , R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
4
From page :
117
To page :
120
Abstract :
A markerless motion tracking (MT) system developed for use in PET brain imaging has been tested in the limited field of view (FOV) of the MR head coil from the Siemens Biograph mMR. The system is a 3D surface scanner that uses structured light (SL) to create point cloud reconstructions of the facial surface. The point clouds are continuously realigned to a reference scan to obtain pose estimates. The system has been tested on a mannequin head performing controlled rotational and translational axial movements within the head coil outside the range of the magnetic field. The RMS of the residual error of the rotation was 0.11° and the RMS difference in the translation with the control system was 0.17 mm, within the trackable range of movement.
Keywords :
Motion tracking , Structured light , PET/MRI , Markerless , Tracoline