• Title of article

    350 keV accelerator based PGNAA setup to detect nitrogen in bulk samples

  • Author/Authors

    Naqvi، نويسنده , , A.A. and Al-Matouq، نويسنده , , Faris A. and Khiari، نويسنده , , F.Z. and Gondal، نويسنده , , M.A. and Rehman، نويسنده , , Khateeb-ur and Isab، نويسنده , , A.A. and Raashid، نويسنده , , M. and Dastageer، نويسنده , , M.A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    6
  • From page
    8
  • To page
    13
  • Abstract
    Nitrogen concentration was measured in explosive and narcotics proxy material, e.g. anthranilic acid, caffeine, melamine, and urea samples, bulk samples through thermal neutron capture reaction using 350 keV accelerator based prompt gamma ray neutron activation (PGNAA) setup. Intensity of 2.52, 3.53–3.68, 4.51, 5.27–5.30 and 10.38 MeV prompt gamma rays of nitrogen from the bulk samples was measured using a cylindrical 100 mm×100 mm (diameter×height ) BGO detector. Inspite of interference of nitrogen gamma rays from bulk samples with capture prompt gamma rays from BGO detector material, an excellent agreement between the experimental and calculated yields of nitrogen gamma rays has been obtained. This is an indication of the excellent performance of the PGNAA setup for detection of nitrogen in bulk samples.
  • Keywords
    Thermal neutron capture based PGNAA setup using 350  , Explosive and narcotic proxy material samples , keV accelerator , BGO detector , Nitrogen detection in bulk samples , Monte Carlo simulations
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2013
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2194625