Title of article :
Calibration procedures to study molecular fragmentation by electron–ion coincidences
Author/Authors :
Thomas Kerkau، نويسنده , , T. and Krنssig، نويسنده , , B. and Schwarzkopf، نويسنده , , O. and Kossmann، نويسنده , , H. and Kنmmerling، نويسنده , , B. and Schmidt، نويسنده , , V.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
13
From page :
555
To page :
567
Abstract :
Calibration procedures are presented for an apparatus in which an electron spectrometer is combined with an ion time-of-flight spectrometer. We analyze the properties of the system for fragmentation studies of doubly-charged diatomic molecules where Auger electrons and ionic fragments are detected in coincidence. Special emphasis is given to the extraction of fragment ion yields from the experimental data and to the measurement of the kinetic energy release. The influence of instrumental parameters and of possible angular distribution effects on the results of such measurements is discussed.
Keywords :
Electron–ion coincidences , Doubly-charged molecule fragmentation
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2002
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2194742
Link To Document :
بازگشت