• Title of article

    Sensitive method for the determination of 18F attached to aerosol particles in a PET centre

  • Author/Authors

    Papp، نويسنده , , Z and Uray، نويسنده , , I، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    9
  • From page
    788
  • To page
    796
  • Abstract
    A new grab sampling method has been developed for the measurement of 18F attached to aerosol particles. It is based on direct β-counting of filtered aerosol sample over successive time intervals by an end-window Geiger–Müller counter. The effect of the progeny of radon and thoron on the β-counting rate is separated by analysing the decay curve. The defined solid angle absolute counting was used to evaluate the efficiencies for 18F and for the progeny of radon and thoron one by one. Absolute activity concentration of 18F can be determined with less than 10% systematic error. Glass-fibre filter and high sampling flow rate are applied, leading to a detection limit for 18F of less than 1 Bq m−3. The method was tested under different circumstances in the PET centre of University of Debrecen, Hungary.
  • Keywords
    radon progeny , GM-counter , 18F , Positron emission tomography , Absolute ?-counting
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2002
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2195003