Title of article :
Upgrade of the liquid xenon detector in the MEG experiment
Author/Authors :
Kaneko، نويسنده , , Daisuke، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
The MEG experiment is searching for the lepton flavor violating decay, μ + → e + + γ , with the highest sensitivity. A major upgrade of the experiment is planned where the performance of the liquid xenon (LXe) γ - ray detector is expected to be significantly improved. The current 2-in. photomultiplier tubes (PMT) on the γ - ray incident face will be replaced with smaller photo-sensors such as multi-pixel photon counters (MPPC) in order to improve the resolutions and the efficiency. A UV-sensitive MPPC operational in LXe is under development in collaboration with Hamamatsu Photonics. A high photon detection efficiency (PDE) of 17% to LXe scintillation light is achieved with a prototype with an active area of 12×12 mm2 and a single-photoelectron peak is clearly observed. The engineering design of the detector and the development of the related technologies such as the signal transmission and assembly are also steadily in progress.
Keywords :
Liquid xenon , MEG experiment , MPPC (SiPM)
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A