Title of article :
Measurement of the γ-ray energy resolution function of EJ301 liquid scintillator using a dual channel ADC
Author/Authors :
Cecconello، نويسنده , , M. and Donato، نويسنده , , M. and Marini-Bettolo، نويسنده , , C. and Conroy، نويسنده , , S. and Sangaroon، نويسنده , , S. and Ericsson، نويسنده , , G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Pages :
4
From page :
34
To page :
37
Abstract :
A comparison of three different methods for the energy calibration of liquid scintillators using γ-ray sources using only a delay unit and a dual channel ADC is presented. Single Compton edge measurements combined with MCNP simulations of the pulse height spectra are compared with the results obtained using Compton coincidence methods. Good agreement between the three methods is found. Measurements from the three different methods are then combined to provide the best estimate of the energy resolution function. The technique here presented could be easily used in laboratory courses as high performing ADCs become more common and affordable.
Keywords :
Liquid scintillator , Coincidence technique , Compton edge , Energy resolution function