Title of article
Analysis of the response of silicon detectors to α particles and 16O ions
Author/Authors
Kirsebom، نويسنده , , O.S. and Fynbo، نويسنده , , H.O.U. and Riisager، نويسنده , , K. and Raabe، نويسنده , , R. and Roger، نويسنده , , T.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2014
Pages
5
From page
57
To page
61
Abstract
We clarify basic concepts concerning the response of silicon detectors to ionizing radiation, focusing on light ions with energies ranging from a few hundred keV to a few MeV. Through reanalysis of existing experimental data, we quantify the difference in the response of silicon detectors to α particles and 16O ions, and we determine the correction that must be made before an energy calibration obtained with α particles can be applied to the measurement of 16O ions. Finally, we show that an apparent disagreement among three recent measurements of the decay spectrum of 8B can be explained by a failure to correctly take into account the different response of silicon detectors to α particles and 16O ions.
Keywords
Charged-particle spectroscopy , Solid-state detectors , Solar neutrinos
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2014
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2196159
Link To Document