Title of article :
Multi-concentric-ring open-air ionization chamber for high-intensity X-ray beams
Author/Authors :
Nariyama، نويسنده , , Nobuteru، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
An ionization chamber with four concentric ring electrodes was used to measure doses of white, 10, 15 and 20 keV synchrotron X-ray beams. The ring-shaped electrodes, which had diameters less than 11.8 mm, collected charges independently only around the beam, excluding strong in-beam charges when the beams passed through a small hole in the electrode centers. As a result, under low saturation voltages, the measured dose rates were confirmed to correlate with the beam intensity when conversion factors calculated with a Monte Carlo code were employed. The influence of the assumed beam sizes and incident positions on the current was almost negligible, with the exception of the incident position dependence at 10 keV.
Keywords :
Ring electrode , Synchrotron radiation , High intensity , ionization chamber
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A