Title of article :
Calibration of LiBaF3:Ce scintillator for fission spectrum neutrons
Author/Authors :
Reeder، نويسنده , , P.L and Bowyer، نويسنده , , S.M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
17
From page :
469
To page :
485
Abstract :
The scintillator LiBaF3 doped with small amounts of Ce+3 has the ability to distinguish heavy charged particles (p, d, t, or α) from beta and/or gamma radiation based on the presence or absence of nanosecond components in the scintillation light output. Since the neutron capture reaction on 6Li produces recoil alphas and tritons, this scintillator also discriminates between neutron induced events and beta or gamma interactions. An experimental technique using a time-tagged 252Cf source has been used to measure the efficiency of this scintillator for neutron capture, the calibration of neutron capture pulse height, and the pulse height resolution—all as a function of incident neutron energy.
Keywords :
Neutron calibration , LiBaF3:Ce scintillator , Time-tagged 252Cf
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2002
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2196277
Link To Document :
بازگشت