Title of article :
Development of a scintillator detector set with counter and data acquisition for flow measurements
Author/Authors :
da Costa، نويسنده , , Fلbio E and Hamada، نويسنده , , Margarida M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
6
From page :
150
To page :
155
Abstract :
A portable counter with data acquisition system for flow measurements was developed, using the pulse velocity technique. This consists in determining the tracer transit time mixed homogeneously to the liquid or gas pipelines. The counter comprises: (a) two CsI(Tl) crystals solid state detectors, associated with Si PIN photodiodes, with compatible sensitivity to the injected radiotracers activities; (b) amplification units; (c) analogue-to-digital interface, which processes and displays the detectors counting separately and in real time, but in a same temporal axis, via a computer screen and (d) 30-m coaxial cables for signals transmission from each detector to the processing unit. Experiments were carried out for the detector and associated electronic characterizations. The equipment showed to be suitable for flow measurements in an industrial plant, in the real situation.
Keywords :
Flux measurement , radiotracer , Transit pulse , scintillator , radiation detector
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2002
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2196562