Abstract :
Recent theoretical damage material models describe the dynamic development of voids and microcracks in materials under plastic deformation. For these models, experimental verification is needed. We propose direct and non-destructive observation of the propagation of material damage by measuring changes in transmission of X-rays penetrating a stressed material, using a photon-counting X-ray imager. The present contribution aims to demonstrate the applicability of silicon and gallium-arsenide devices for X-ray transmission measurements with a specimen of high-ductile aluminium alloy under study. The first experiments to determine the resolution and the sensitivity of the proposed method with the Medipix-1 pixel detector are presented.
Keywords :
X- and gamma-ray detectors , Imaging detectors , digital radiography , Defectoscopy , Fracture mechanics , Damage , Voids , Semiconductor pixel detectors , Position- sensitive detectors , X-ray and gamma-ray measurements , X-ray radiography