Title of article :
Investigation on gamma-ray position sensitivity at 662 keV in a spectroscopic 3” x 3” LaBr3:Ce scintillator
Author/Authors :
Giaz، نويسنده , , A. and Camera-Roda، نويسنده , , F. and Birocchi، نويسنده , , F. and Blasi، نويسنده , , N. and Boiano، نويسنده , , C. and Brambilla، نويسنده , , S. and Coelli، نويسنده , , S. and Fiorini، نويسنده , , C. and Marone، نويسنده , , A. and Million، نويسنده , , B. and Riboldi، نويسنده , , S. and Wieland، نويسنده , , O.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2015
Pages :
9
From page :
103
To page :
111
Abstract :
The position sensitivity of a thick, cylindrical and continuous 3” x 3” (7.62 cm x 7.62 cm) LaBr3:Ce crystal was studied using a 1 mm collimated beam of 662 keV gamma rays from a 400 MBq intense 137Cs source and a spectroscopic photomultiplier (PMT) (HAMAMATSU R6233-100SEL). The PMT entrance window was covered by black absorber except for a small window 1 cm x 1 cm wide. lete scan of the detector over a 0.5 cm step grid was performed for three positions of the 1 cm x 1 cm window. For each configuration the energy spectrum was measured and the peak centroid, the FWHM, the area and peak asymmetry of the 662 keV gamma transition were analyzed. The data show that, even in a 3” thick LaBr3:Ce crystal with diffusive surfaces the position of the full energy peak centroid depends on the source position. We verified that, on average, the position of the full energy peak centroids measured in the three 1 cm x 1 cm window configurations is sufficient for the correct identification of the collimated gamma source position.
Keywords :
Position sensitivity , Doppler broadening , Imaging , LaBr3:Ce scintillators
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2015
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2196977
Link To Document :
بازگشت