    The efficiency calibration of a gamma spectroscopy waste assay system, constructed by IFIN-HH, was performed. The calibration technique was based on the assumption of a uniform distribution of the source activity in the drum and also a uniform sample matrix. A collimated detector (HPGe—20% relative efficiency) placed at 30 cm from the drum was used. The detection limit for 137Cs and 60Co is approximately 45 Bq/kg for a sample of about 400 kg and a counting time of 10 min. A total measurement uncertainty of −70% to +40% was estimated.