Title of article :
Comment on “Perspectives of medical X-ray imaging”
Author/Authors :
Taibi، نويسنده , , A and Baldelli، نويسنده , , P and Tuffanelli، نويسنده , , A and Gambaccini، نويسنده , , M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
3
From page :
685
To page :
687
Abstract :
In the paper “Perspectives of medical X-ray imaging” (Nucl. Instr. and Meth. A 466 (2001) 99) the authors infer, from simple approximations, that the use of HOPG monochromator has no advantage in mammography compared to existing systems. We show that in order to compare imaging properties of different X-ray sources it is necessary to evaluate the spectra after the attenuation of the tissue to be imaged. Indeed, quasi-monochromatic X-ray sources have the potential to enhance image contrast and to reduce patient dose.
Keywords :
mammography , Quasi-monochromatic X-rays , X-ray sources , Mosaic crystals