Title of article :
A comprehensive calibration method of Ge detectors for low-level gamma-spectrometry measurements
Author/Authors :
Pérez-Moreno، نويسنده , , J.P. and San Miguel، نويسنده , , E.G. and Bol??var، نويسنده , , J.P. and Aguado، نويسنده , , J.L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
11
From page :
152
To page :
162
Abstract :
This paper aims to show a general and simple method for gamma-ray efficiency calibration in the 186–1500 keV energy range with special attention to applications on environmental studies. This method accounts for differences in densities, composition and counting geometries between real and calibration samples. Self-absorption corrections can be easily determined from a single spectral measurement knowing only the apparent densities of the samples. We have evaluated the influence of background parameters in the response of the spectrometric system and moreover, we have calculated the minimum detectable activity due to background contribution. This calibration has been validated through three different and independent tests.
Keywords :
Low-level counting , Gamma-spectrometry , efficiency calibration , Self-absorption , Ge detectors
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2002
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2197300
Link To Document :
بازگشت