• Title of article

    Beam energy calibration for the folded tandem ion accelerator

  • Author/Authors

    Santra، نويسنده , , S. and Mahata، نويسنده , , K. and Singh، نويسنده , , P. and Fernandes، نويسنده , , C.V. and Hemalatha، نويسنده , , M. and Kailas، نويسنده , , S.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    44
  • To page
    50
  • Abstract
    A simple method, based on Back Scattering (BS) measurements, has been used to calibrate the analysing magnet of the Folded Tandem Ion Accelerator (FOTIA). The combination of various beams with different charge states and at different terminal voltages covering a range of magnetic fields has yielded the calibration constant ‘K’ of the magnet with an estimated error of about ±0.1%. The resonance in 12C(p,p)12C at Elabp=4.808 MeV (the excited state, 5/2+, of 13N with a width of 11 keV) has also been measured. The ‘K’ value determined from this is consistent with that obtained using BS.
  • Keywords
    Voltage stability , Folded tandem ion accelerator , Analysing magnet , Energy calibration , Back scattering , Resonance reaction
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2003
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number