Title of article :
Energy degrader optimization for medical beam lines
Author/Authors :
Anferov، نويسنده , , Vladimir، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
6
From page :
222
To page :
227
Abstract :
This paper describes the optimization of a variable energy degrader design for the Midwest Proton Radiotherapy Institute (MPRI) (The Indiana University Midwest Proton Radiation Institute, in: Proceedings of the 2001 Particle Accelerator Conference, PAC-2001, Chicago, 2002, p. 645). To optimize the energy degrader design we investigate the choice of an optimal material for the degrader, the beam emittance growth in the degrader, and the matching of the degraded beam with the acceptance of a medical beam line.
Keywords :
Medical beam lines , Radiotherapy equipment , Energy degrader