Title of article :
A systematic study of large PMTs for the Pierre Auger Observatory
Author/Authors :
Tripathi، نويسنده , , A.K. and Akhanjee، نويسنده , , S. and Arisaka، نويسنده , , K. and Barnhill، نويسنده , , D. and DʹPasquale، نويسنده , , C. and Jillings، نويسنده , , C. and Ohnuki، نويسنده , , T. and Ranin، نويسنده , , P.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
9
From page :
331
To page :
339
Abstract :
In order to select the most suitable Photomultiplier Tubes (PMTs) for the surface detectors of the Pierre Auger Observatory, we have performed extensive tests on large PMTs from three different manufacturers: Electron Tubes Limited, Hamamatsu Photonics, and Photonis Imaging Sensors. We have developed specialized instrumentation to accurately measure the PMT characteristics such as gain, dark current, single photoelectron spectrum, dark pulse rate, afterpulse ratio, quantum efficiency, and linearity. The instrumentation and techniques used for these measurements and their results are reported.
Keywords :
PMTs , Water cherenkov , Cosmic rays , Astroparticle physics , Photodetectors
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2003
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2197854
Link To Document :
بازگشت