Title of article :
Two-dimensional GEM imaging detector with delay-line readout
Author/Authors :
Guedes، نويسنده , , G.P. and Breskin، نويسنده , , A and Chechik، نويسنده , , R and Vartsky، نويسنده , , D and Bar، نويسنده , , D and Barbosa، نويسنده , , A.F. and Marinho-Soriano، نويسنده , , P.R.B، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
11
From page :
473
To page :
483
Abstract :
A 100×100 mm2 2D imaging detector, based on a triple-GEM gaseous multiplier, striped x−y readout anode and discrete delay-line readout, is presented. The fast (2.1 ns tap−1) delay-line circuit was designed to match the anode-charge signal profile, namely its rise-time and length. The detectorʹs imaging capability was systematically studied in Ar/CO2 (70/30) with 5.9 keV X-rays; x−y resolution of σ=0.05 and 0.1 mm for top and bottom anode strips, respectively, and integral non-linearity of ∼0.15% are demonstrated.
Keywords :
2D readout , GEM , Gaseous imaging detector , Delay-line readout