• Title of article

    CsI(Tl)/plastic phoswich detector enhanced in low-energy gamma-ray detection

  • Author/Authors

    K. Yamasoto، نويسنده , , Koutaro and Tsutsumi، نويسنده , , Masahiro and Oishi، نويسنده , , Tetsuya and Yoshizawa، نويسنده , , Michio and Yoshida، نويسنده , , Makoto، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    609
  • To page
    615
  • Abstract
    A phoswich detector composed of a thin plate CsI(Tl) scintillator and a plastic scintillator (BC-400) has been designed and evaluated in order to improve the sensitivity in the low-energy region of a large-area plastic scintillation detector. This newly designed phoswich detector can be applied to both gross gamma measurement and energy spectrometry for low-energy gamma-ray emitters. Judging by estimations of minimum detectable activity, the lower measurable energy of a large-area plastic scintillation detector can be expanded down to a few tens of keV by adding a thin plate CsI(Tl) scintillator.
  • Keywords
    Low-energy region , Thin plate CsI(Tl) scintillator , Large-area plastic scintillation detector , Rise-time discrimination technique , Phoswich detector , minimum detectable activity
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2005
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2199224