Title of article :
Portable, low-power, mechanically cooled Ge spectrometer
Author/Authors :
Becker، نويسنده , , J.A. and Cork، نويسنده , , C.P. and Fabris، نويسنده , , L. and Madden، نويسنده , , N.W.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
3
From page :
167
To page :
169
Abstract :
A light-weight portable mechanically cooled γ-ray spectrometer has been constructed and tested. The spectrometer is based on a high-purity n-type Ge coaxial crystal, ∼5 cm long×5 cm diameter, a small, low-power Stirling cycle microcooler, and a low-power custom electronics package. The energy resolution of the spectrometer is ∼3.5 keV at Eγ=662 keV, the power requirements are ∼15 W DC, and the combined weight of the Ge, housing, and controller is approximately 10 pounds. The spectrometer qualifies therefore as “hand held”. It is suitable for field operations, because of its light-weight, low-power draw, and operational lifetime. The microcooler itself has a MTBF >30,000 h, and the spectrometer runs for several months (at least 6) before a 2 day recycle is required.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2003
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2199247