• Title of article

    Comparison of 5 and 10 mm thick HgI2 pixelated γ-ray spectrometers

  • Author/Authors

    Baciak، نويسنده , , James E. and He، نويسنده , , Zhong، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    4
  • From page
    191
  • To page
    194
  • Abstract
    HgI2 detectors fabricated with pixelated anodes were tested as potential room temperature spectrometers. Spectra were obtained with 5 and 10 mm thick devices, and their results were compared. The 5 mm thick devices showed some effect of the anode weighting potential, while the 10 mm thick devices showed the effects of charge diffusion/sharing and electron trapping. A resolution near 1.4% FWHM using a 137Cs source was achieved after correcting for the interaction depth within the detector. The resolution from individual depths was as low as approximately 1.1% FWHM.
  • Keywords
    Mercuric iodide , Compound Semiconductor , Room temperature , ?-ray detectors
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2003
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number