Title of article :
Performance study of monolithic line radiation detector based on semi-insulating GaAs using X-ray source
Author/Authors :
Za?ko، نويسنده , , B. and Dubeck?، نويسنده , , F. and ??epko، نويسنده , , P. and Melov، نويسنده , , V. and Herms، نويسنده , , M. A. Haupt، نويسنده , , L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
The 32-strips linear detector based on semi-insulating GaAs is fabricated. The dimensions of each strip are 170×2500×200 μm3. Analog and digital read-out electronic card with 24 channels is connected to the linear detector. The surface mounted device (SMD) components with automatic assembling of electronic devices for read-out front-end electronics are used. Testing of counting ability of each strip with micro-collimated (φ∼250 μm) X-ray source up to 60 kV is carried out. The maximum counting rate is about 1.0×105 cps. The results for gradual decrease of linearity on approaching the maximum counting rate and the dispersion of the intensity registered in individual channels are presented. The crosstalk between neighbouring strips is found negligible.
Keywords :
digital imaging , Strip line detector , X-ray irradiation , Semi-insulating GaAs
Journal title :
Nuclear Instruments and Methods in Physics Research Section A