Title of article :
Study of the response function of a HPGe detector for low-energy X-rays
Author/Authors :
Lépy، نويسنده , , Marie-Christine and Plagnard، نويسنده , , Johann and Ferreux، نويسنده , , Laurent، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
4
From page :
290
To page :
293
Abstract :
Accurate processing of low-energy X-ray spectra requires detailed knowledge of the detector response function. This has been determined for a high-purity germanium (HPGe) detector by using monochromatized synchrotron radiation in the 1–20 keV energy range. Due to the high peak-to-background ratio, the spectra reveal shapes that were hidden using previous detectors. In addition to escape peaks, features caused by electron interactions characterize the background shape: escape of photoelectrons and Auger electrons from the active part of the crystal and penetration of the electrode ones into the detector. The tunable incident radiation makes it possible to examine and identify these different features by fine scanning at the binding energies of the detector materials.
Keywords :
escape , X-ray spectra , Low-energy X-rays
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2003
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2199295
Link To Document :
بازگشت