Title of article :
The dependence of the virtual point-detector on the HPGe detector dimensions
Author/Authors :
Mahling، نويسنده , , S. and Orion، نويسنده , , I. and Alfassi، نويسنده , , Z.B.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
10
From page :
544
To page :
553
Abstract :
The dependence of the HPGe detector efficiency when collecting full-energy peaks vs. the distance of the point source from the detector cap was studied by Monte-Carlo methods for 49 cylindrical detectors, having different radii and thicknesses. It was shown that in all cases, interpolation and extrapolation can be done using the model of the virtual point-detector. The dependence of the distance of the virtual point-detector position h0 from the detector cap was studied as a function of the geometrical dimension of the detector. Assigning r and h to the radius and the thickness of the detector, a, b, c to parameters depending on the photon energy and μ/ρ to the mass attenuation coefficient, a general formula was found: h 0 = ar ( 1 - b exp [ - hc μ / ρ ] ) . The parameter c was found to change very little with the photon energy (±1%), while a and b depend more strongly on the photon energy (±5%).
Keywords :
Virtual point-detector , Detector radius , Detector length , Detector efficiency calibration , MCNP , Monte-Carlo simulation
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2006
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2199700