Title of article :
Aging in large CDF tracking chambers
Author/Authors :
Binkley، نويسنده , , M. and Wagner، نويسنده , , R.L. and Mukherjee، نويسنده , , Pedro A. and Ambrose، نويسنده , , D. and Bauer، نويسنده , , G. and Khazins، نويسنده , , D.M. and Atac، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
The experience of the Collider Detector at Fermilab (CDF) with aging in the large axial drift chamber responsible for tracking in the central region is presented. Premature aging in the Run 1 chamber was observed after only 0.02 C/cm. After cleaning much of the gas system and making modifications to reduce aerosols from the alcohol bubbler, the observed aging rate fell dramatically in test chambers. Considerable effort has been made to better understand the factors that affect aging since the replacement chamber for Run 2 will accumulate about 1.0 C/cm. Current test chambers using the full CDF gas system show aging rates of less than 5%/C/cm.
Keywords :
chamber , wire , CLEAN , gaseous , argon , Ethane , Aging , alcohol
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A