Title of article
Preliminary results from a Monte Carlo study of breast tumour imaging with low-energy high-resolution collimator and a modified uniformly-redundant array-coded aperture
Author/Authors
Alnafea، نويسنده , , M. and Wells، نويسنده , , K. and Spyrou، نويسنده , , N.M. and Saripan، نويسنده , , M.I and Guy، نويسنده , , M. J. Hinton، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
4
From page
146
To page
149
Abstract
This paper considers scintimammography imaging using two different image acquisition/formation methods: Low-Energy High-Resolution (LEHR) collimator, and Modified Uniformly Redundant Arrays (MURAs)-coded aperture (CA) systems. Using Monte-Carlo simulation, the effectiveness and performance of both systems were evaluated by quantitative comparison of lesion contrast and full-width half-maximum under a variety of clinical imaging situations. The results from the CA studies suggest that for the particular geometry studied, the MURAs-CA gives comparable performance compared to LEHR collimator.
Keywords
scintimammography , Breast tumuor imaging , Coded aperture imaging , Mura , Monte-Carlo simulation , Phantom study
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2006
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2200957
Link To Document