Title of article :
Experimental system for high resolution X-ray transmission radiography
Author/Authors :
Jakubek، نويسنده , , Jan and Holy، نويسنده , , Tomas and Jakubek، نويسنده , , Martin and Vavrik، نويسنده , , Daniel and Vykydal، نويسنده , , Zdenek، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
The system presented provides an experimental setup with stable position of both X-ray tube and detector as well as the operational movement of the observed object fixed on a motorized stage. Linear and rotation axes of the computer controlled stage enable radiographic scanning of objects as well as their tomographic reconstruction. High quality radiographs are obtained even for very low contrast and low absorption objects thanks to the large dynamic range of the cooled X-ray imager together with the proposed beam hardening correction procedure. This procedure is fully automated using a set of calibration filters and appropriate software modules.
Keywords :
digital radiography , Roentgenogram quality , Beam hardening , Single photon counting pixel device
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A