• Title of article

    Readout system and predicted performance of a low-noise low-frequency horizontal accelerometer

  • Author/Authors

    Bertolini، نويسنده , , Alessandro and DeSalvo، نويسنده , , Riccardo and Fidecaro، نويسنده , , Francesco and Francesconi، نويسنده , , Mario and Marka، نويسنده , , Szabolcs and Sannibale، نويسنده , , Virginio and Simonetti، نويسنده , , Duccio and Takamori، نويسنده , , Akiteru and Tariq، نويسنده , , Hareem، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    8
  • From page
    579
  • To page
    586
  • Abstract
    The readout system of a new very low noise low-frequency, force-feedback accelerometer is presented. The horizontal accelerometer carrying this sensor has been designed to be integrated in an advanced seismic isolation system for interferometric gravitational wave detectors. To make it suitable for ultra-high-vacuum operation and to reduce the probability of failures of the in-vacuum components, the accelerometer is equipped with a high-resolution capacitance sensor fully driven by remote signal conditioning electronics. The instrumentʹs performance is not compromised by cabling lengths of up to 15 m. The accelerometer, equipped with this readout scheme, should be able to detect the inertial displacement of a platform with a root-mean-square noise less than 1 nm, integrated over all the frequencies above 0.01 Hz. We plan to implement the same readout scheme in a companion vertical sensor now being designed.
  • Keywords
    flexure , Capacitance sensor , Accelerometer , Matching transformer
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2006
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number