An alternative approach for the determination of the full-peak detection efficiency in case of a close-in detection geometry
Author/Authors
De Wispelaere، نويسنده , , A. and De Corte، نويسنده , , F.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
5
From page
645
To page
649
Abstract
The aim of this work is to propose an alternative method for the determination of εp in the case of measurements restricted to one single sample-detector distance. It is based on an iterative procedure, linking the measurement of calibrated multi-γ point sources at this particular detector distance (the resulting εp-function being disturbed by true-coincidence effects) with measurement of some coincidence-free single-γ point sources (leading to peak-to-total ratios P/T that, combined with detection efficiencies, yield correction factors for true-coincidence effects). The thus obtained εp versus Eγ curve for point-source geometry can then be converted to the actual extended sample geometry, based on the effective solid angle concept.
Keywords
Activation analysis , Detection efficiency , ?-ray spectrometry