Title of article :
Cryogenic neutron detector by InSb semiconductor detector with high-density helium-3 gas converter
Author/Authors :
Nakamura، نويسنده , , T and Katagiri، نويسنده , , M and Aratono، نويسنده , , Y and Kanno، نويسنده , , I and Hishiki، نويسنده , , S and Sugiura، نويسنده , , O and Murase، نويسنده , , Y، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
4
From page :
76
To page :
79
Abstract :
The neutron-detection characteristics of a cryogenic neutron detector comprising an InSb semiconductor detector and a helium-3 gas converter were evaluated at a gas pressure up to 12.4 atm at 4.2 K. The detector successfully detected stable neutrons under these conditions, where the density of the helium-3 gas is a few-hundred times higher than that at room temperature. It was found that the neutron detection efficiency was correlated with the gas pressure—even in a backward-detection configuration—in low-temperature, high-pressure helium-3.
Keywords :
InSb , Helium-3 , Semiconductor detector , High spatial resolution , neutron detector
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2004
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2201555